i Oxygen Analyzer Teledyne Analytical InstrumentsOPERATING INSTRUCTIONS FOR Model 3010TAC Trace Oxygen AnalyzerFlush Mount Control Unit, PN D-
ii: Part IModel 3010TAC Oxygen Analyzer Teledyne Analytical InstrumentsTable of Contents1 Introduction1.1 Overview ...
Part I: iii Part I: Control Unit Teledyne Analytical Instruments4.3.6 Version Screen ... 4-104.3.7
iv: Part IModel 3010TAC Oxygen Analyzer Teledyne Analytical Instruments
Part I: 1-1Oxygen Analyzer Part I: Control Unit Teledyne Analytical InstrumentsIntroduction1.1 OverviewThe Analytical Instruments, Remote Probe and I
1-2: Part I1 Introduction Model 3010TAC Teledyne Analytical InstrumentsFigure 1-1: Front of Unmounted Control UnitFunction Keys: Six touch-sensit
Part I: 1-3Oxygen Analyzer Part I: Control Unit Teledyne Analytical Instruments• Enter Moves VFD display on to the next screen in a series.If none re
1-4: Part I1 Introduction Model 3010TAC Teledyne Analytical InstrumentsFigure 1-2: Model 3010TAC Rear Panel• Power Connection Universal AC power so
Part I: 1-5Oxygen Analyzer Part I: Control Unit Teledyne Analytical Instruments• Remote Probe Receives remote probe signals• Network I/O Serial digit
1-6: Part I1 Introduction Model 3010TAC Teledyne Analytical Instruments
Part I: 2-1Oxygen Analyzer Part I: Control Unit Teledyne Analytical InstrumentsOperational Theory2.1 IntroductionThe Model 3010TAC Oxygen Analyzer Co
iiModel 3010TAC Teledyne Analytical InstrumentsCopyright © 1999 Teledyne Analytical InstrumentsAll Rights Reserved. No part of this manual may be repr
2-2: Part I2 Operational Theory Model 3010TAC Teledyne Analytical InstrumentsFigure 2-1: Block Diagram of the 3010TAC
Part I: 2-3Oxygen Analyzer Part I: Control Unit Teledyne Analytical InstrumentsThe digital concentration signal—along with input from the controlpane
2-4: Part I2 Operational Theory Model 3010TAC Teledyne Analytical Instruments
Oxygen Analyzer Part I: Control UnitPart I: 3-1 Teledyne Analytical InstrumentsInstallationInstallation of Model 3010TAC Analyzers includes:1. Unpack
3 Installation Model 3010TAC3-2: Part I Teledyne Analytical InstrumentsFigure 3-1: Front Panel of the Model 3010TAC Control UnitFigure 3-2: Single
Oxygen Analyzer Part I: Control UnitPart I: 3-3 Teledyne Analytical Instrumentsis pressed all the way in with a narrow gauge tool (less than 0.18 inc
3 Installation Model 3010TAC3-4: Part I Teledyne Analytical InstrumentsCAUTION: Use Shielded Cables. Also, use plugs that provideexcellent EMI/RFI p
Oxygen Analyzer Part I: Control UnitPart I: 3-5 Teledyne Analytical InstrumentsAnalog Outputs: There are four DC output signal pins—two pinsper outp
3 Installation Model 3010TAC3-6: Part I Teledyne Analytical Instruments• Can be configured as failsafe or nonfailsafe.• Can be configured as latch
Oxygen Analyzer Part I: Control UnitPart I: 3-7 Teledyne Analytical Instrumentssynchronous signal must open and close external span valveappropriatel
iii Oxygen Analyzer Teledyne Analytical InstrumentsTable of ContentsSpecific Model Information... ivPreface ...
3 Installation Model 3010TAC3-8: Part I Teledyne Analytical Instrumentssignals to ensure that the zero and span gas valves will becontrolled synchro
Oxygen Analyzer Part I: Control UnitPart I: 3-9 Teledyne Analytical InstrumentsThe voltage from these outputs is nominally 0 V for the OFF and15 V dc
3 Installation Model 3010TAC3-10: Part I Teledyne Analytical InstrumentsTable 3-5: Commands via RS-232 InputCommand Descriptionas<enter> Immed
Oxygen Analyzer Part I: Control UnitPart I: 3-11 Teledyne Analytical InstrumentsCAUTION: If you use your own control valves, or any otherdesign, plea
3 Installation Model 3010TAC3-12: Part I Teledyne Analytical Instruments
Oxygen Analyzer Part I: Control UnitPart I: 4-1 Teledyne Analytical InstrumentsOperation4.1 IntroductionOnce the analyzer has been installed, configu
4 Operation Model 3010TAC4-2: Part I Teledyne Analytical Instruments4.2 Using the Data Entry and FunctionButtonsData Entry Buttons: The < > ar
Oxygen Analyzer Part I: Control UnitPart I: 4-3 Teledyne Analytical Instruments4.3 The System FunctionThe subfunctions of the System function are de
4 Operation Model 3010TAC4-4: Part I Teledyne Analytical InstrumentsAfter a password is assigned, the operator must log out toactivate it. Until the
Oxygen Analyzer Part I: Control UnitPart I: 4-5 Teledyne Analytical InstrumentsIn the first line, TRACK or HOLD should be blinking. The operatorcan t
ivModel 3010TAC Teledyne Analytical InstrumentsSpecific Model InformationThe instrument for which this manual was supplied may incorporate oneor more
4 Operation Model 3010TAC4-6: Part I Teledyne Analytical InstrumentsSpan OFF Nxt: 0d 0hZero OFF Nxt: 0d 0hPress < > arrows to blink Span (or
Oxygen Analyzer Part I: Control UnitPart I: 4-7 Teledyne Analytical InstrumentsTRAK/HLD Auto—CalPSWD Logout MoreUse the < > arrow keys to scrol
4 Operation Model 3010TAC4-8: Part I Teledyne Analytical InstrumentsIf you chose Enter to change the password, the password assignmentscreen appears
Oxygen Analyzer Part I: Control UnitPart I: 4-9 Teledyne Analytical Instruments0.0 ppm AnlZRange: 0 — 100If an alarm is tripped, the second
4 Operation Model 3010TAC4-10: Part I Teledyne Analytical InstrumentsTRAK/HLD Auto—CalPSWD Logout MoreUse the < > arrow keys to blink More, th
Oxygen Analyzer Part I: Control UnitPart I: 4-11 Teledyne Analytical Instruments- Use the Right or Left arrow keys and select More. Press Enter.Vers
4 Operation Model 3010TAC4-12: Part I Teledyne Analytical InstrumentsIf you are using password protection, you will need to enter yourpassword to ga
Oxygen Analyzer Part I: Control UnitPart I: 4-13 Teledyne Analytical Instruments4.4.1.2 Manual Mode ZeroingPress Zero to enter the Zero function. The
4 Operation Model 3010TAC4-14: Part I Teledyne Analytical Instruments4.4.2 Span CalThe Span button on the front panel is used to span calibrate the
Oxygen Analyzer Part I: Control UnitPart I: 4-15 Teledyne Analytical Instrumentssponds, within tolerance, to the value of the span gas concentration.
v Oxygen Analyzer Teledyne Analytical InstrumentsPrefaceOverviewThe Teledyne Analytical Instruments Model 3010TAC Trace OxygenAnalyzer is a versat
4 Operation Model 3010TAC4-16: Part I Teledyne Analytical Instrumentsthe full scale of the range being calibrated for a period of ten minutes it iss
Oxygen Analyzer Part I: Control UnitPart I: 4-17 Teledyne Analytical InstrumentsDecide how your alarms should be configured. The choice will dependup
4 Operation Model 3010TAC4-18: Part I Teledyne Analytical InstrumentsSet up alarm 1 by moving the blinking over to AL–1 using the < >arrow key
Oxygen Analyzer Part I: Control UnitPart I: 4-19 Teledyne Analytical Instrumentsrange limits are exceeded, the instrument will automatically shift to
4 Operation Model 3010TAC4-20: Part I Teledyne Analytical InstrumentsTo switch from autoranging to fixed range analysis, enter the rangefunction by
Oxygen Analyzer Part I: Control UnitPart I: 4-21 Teledyne Analytical Instrumentsconcentration and one range ID) and two isolated 4-20 mA dc currentou
4 Operation Model 3010TAC4-22: Part I Teledyne Analytical InstrumentsIMPORTANT:In the event of loss of flow through the analyzer, if thevent is vent
Part I: Control Unit Maintenance 5Part I: 5-1 Teledyne Analytical InstrumentsMaintenanceAside from normal cleaning and checking for leaks at the ga
5 Maintenance Model 3010TAC Oxygen Analyzer5-2: Part I Teledyne Analytical Instruments3. Replace fuse as shown in Figure 5-2.4. Reassemble Housing
Part I: Control Unit Maintenance 5Part I: 5-3 Teledyne Analytical Instruments5.3 Major Internal ComponentsThe major components in the Control Unit
viModel 3010TAC Teledyne Analytical InstrumentsModel and Part Number DesignationsThe part numbers are the most specific identification. When using thi
5 Maintenance Model 3010TAC Oxygen Analyzer5-4: Part I Teledyne Analytical InstrumentsxxxxxFigure 5-4: Rear-Panel ScrewsTo detach the rear panel, r
Part II: i Part II: Analysis Unit Teledyne Analytical InstrumentsOPERATING INSTRUCTIONSModel 3010TACOxygen AnalyzerPart IIIntrinsic Safe Barriersand
ii: Part IIModel 3010TAC Oxygen Analyzer Teledyne Analytical InstrumentsTable of Contents1 Introduction1.1 Overview ...
Part II: 1-1Oxygen Analyzer Part II: Analysis Unit Teledyne Analytical InstrumentsIntroduction1.1 OverviewThe Analytical Instruments Model 3010TAC Re
1-2: Part II1 Introduction Model 3010TAC Teledyne Analytical InstrumentsCAUTION: Bypassing the barriers in any way nullifies theireffect, and cond
Part II: 2-1Oxygen Analyzers Part II: Analysis Units Teledyne Analytical InstrumentsOperational Theory2.1 IntroductionThe Analysis Unit is composed
2-2: Part II2 Operational Theory Model 3010TAC Teledyne Analytical Instrumentsbeing analyzed. The Micro-Fuel Cell is therefore a hybrid between abat
Part II: 2-3Oxygen Analyzers Part II: Analysis Units Teledyne Analytical InstrumentsAt the top end of the cell is a diffusion membrane of Teflon, wh
2-4: Part II2 Operational Theory Model 3010TAC Teledyne Analytical InstrumentsThe overall reaction for the fuel cell is the SUM of the half reaction
Part II: 2-5Oxygen Analyzers Part II: Analysis Units Teledyne Analytical InstrumentsFigure 2-3. Characteristic Input/Output Curve for a Micro-Fuel
vii Oxygen Analyzer Teledyne Analytical InstrumentsMain Features of the AnalyzerThe Model 3010TAC series Oxygen Analyzers are sophisticated yetsim
2-6: Part II2 Operational Theory Model 3010TAC Teledyne Analytical Instruments
Oxygen Analyzer Part II: Analysis UnitPart II: 3-1 Teledyne Analytical InstrumentsInstallationInstallation of the Model 3010TAC Analyzer includes:1.
3 Installation Model 3010TAC3-2: Part II Teledyne Analytical InstrumentsIf you use your own gas control valves, use the interconnect diagramin Figur
Oxygen Analyzer Part II: Analysis UnitPart II: 3-3 Teledyne Analytical Instruments3.5 Intrinsic Safety BarriersTwo dual-channels intrinsic safety bar
3 Installation Model 3010TAC3-4: Part II Teledyne Analytical InstrumentsD - SUB CONNECTOR’S DESCRIPTIONPIN # Description 3 + Range ID 4-20 mA 4 -
Oxygen Analyzer Part II: Analysis UnitPart II: 3-5 Teledyne Analytical Instruments 40 Calibration Contact 41 Calibration Contact 42 Alarm 2 NC Contac
3 Installation Model 3010TAC3-6: Part II Teledyne Analytical Instruments
Oxygen Analyzer Part II: Analysis UnitPart II: 4-1 Teledyne Analytical InstrumentsMaintenance4.1 Routine MaintenanceAside from normal cleaning and ch
4 Maintenance Model 3010TAC4-2: Part II Teledyne Analytical Instruments4.2.1 Installing a New Micro-Fuel CellIt is important to minimize the amou
Oxygen Analyzer Part II: Analysis UnitPart II: 4-3 Teledyne Analytical Instruments4.2.2 Cell WarrantyThe Micro-Fuel cell used in the standard Model
viiiModel 3010TAC Teledyne Analytical InstrumentsModel 3010TAC complies with all of the requirements of theCommonwealth of Europe (CE) for Radio Freq
4 Maintenance Model 3010TAC4-4: Part II Teledyne Analytical Instruments0OK1 DAC A (0–1 V Concentration)2 DAC B (0–1 V Range ID)3 Both FailedPreamp0
A-1Oxygen Analyzer Appendix Teledyne Analytical InstrumentsOPERATING INSTRUCTIONSModels 3010TACOxygen AnalyzersAppendixFlush Mount Control Unit, PN D6
A-2Appendix Model 3010TAC Teledyne Analytical InstrumentsContentsA-1 Model 3010TAC Specifications... A-3A-2 Rec
A-3Oxygen Analyzer Appendix Teledyne Analytical InstrumentsAppendixA-1 Models 3010TAC SpecificationsPackaging: General Purpose Control Unit• Flush pa
A-4Appendix Model 3010TAC Teledyne Analytical InstrumentsPower:General Purpose Control UnitUniversal power supply 85-250 V ac,47-63 Hz.Operating Te
A-5Oxygen Analyzer Appendix Teledyne Analytical InstrumentsA-2 Recommended 2-Year Spare Parts ListQty Part Number Description1 C67821 Back Panel Boa
A-6Appendix Model 3010TAC Teledyne Analytical InstrumentsA-3 Drawing ListD-66192A:Final Assembly/Outline Drawing, Control Unit, Trace OxygenD-66193
A-7Oxygen Analyzer Appendix Teledyne Analytical InstrumentsMeltingPoint(°C)328360A-4 Material Safety Data SheetProduct Name: Micro-Fuel Cells and Supe
A-8Appendix Model 3010TAC Teledyne Analytical InstrumentsFlash Point: na Flammable Limits: na LEL: na UEL: naExtinguishing Media: Use extinguishing
A-9Oxygen Analyzer Appendix Teledyne Analytical InstrumentsContact of electrolyte with skin or eyes will cause aburning sensation and/or feel soapy or
Part I: i Part I: Control Unit Teledyne Analytical InstrumentsOPERATING INSTRUCTIONSModel 3010TACOxygen AnalyzerPart I: Control UnitFlush Panel Mount
A-10Appendix Model 3010TAC Teledyne Analytical InstrumentsSection VIII – Control MeasuresEye Protection: Chemical splash gogglesHand Protection: Ru