OPERATING INSTRUCTIONS FOR
Model 3010TAC
Trace Oxygen Analyzer
Flush Mount Control Unit, PN D-66192A
CENELEC Type Remote Probe, PN C-66336
Intrinsic Safe Barriers Assy., PN C-67564
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS
MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PER-
SIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING.
BEFORE CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED
SUPERVISOR/MANAGER.
DANGER
P/N M66394
11/22/99
ECO#:99-0459
Przeglądanie stron 0
1 2 3 4 5 6 ... 89 90

Podsumowanie treści

Strona 1 - Model 3010TAC

i Oxygen Analyzer Teledyne Analytical InstrumentsOPERATING INSTRUCTIONS FOR Model 3010TAC Trace Oxygen AnalyzerFlush Mount Control Unit, PN D-

Strona 2 - Model 3010TAC

ii: Part IModel 3010TAC Oxygen Analyzer Teledyne Analytical InstrumentsTable of Contents1 Introduction1.1 Overview ...

Strona 3 - Oxygen Analyzer

Part I: iii Part I: Control Unit Teledyne Analytical Instruments4.3.6 Version Screen ... 4-104.3.7

Strona 4 - Specific Model Information

iv: Part IModel 3010TAC Oxygen Analyzer Teledyne Analytical Instruments

Strona 5

Part I: 1-1Oxygen Analyzer Part I: Control Unit Teledyne Analytical InstrumentsIntroduction1.1 OverviewThe Analytical Instruments, Remote Probe and I

Strona 6

1-2: Part I1 Introduction Model 3010TAC Teledyne Analytical InstrumentsFigure 1-1: Front of Unmounted Control UnitFunction Keys: Six touch-sensit

Strona 7 - Main Features of the Analyzer

Part I: 1-3Oxygen Analyzer Part I: Control Unit Teledyne Analytical Instruments• Enter Moves VFD display on to the next screen in a series.If none re

Strona 8

1-4: Part I1 Introduction Model 3010TAC Teledyne Analytical InstrumentsFigure 1-2: Model 3010TAC Rear Panel• Power Connection Universal AC power so

Strona 9

Part I: 1-5Oxygen Analyzer Part I: Control Unit Teledyne Analytical Instruments• Remote Probe Receives remote probe signals• Network I/O Serial digit

Strona 10

1-6: Part I1 Introduction Model 3010TAC Teledyne Analytical Instruments

Strona 11 - Part I: iii

Part I: 2-1Oxygen Analyzer Part I: Control Unit Teledyne Analytical InstrumentsOperational Theory2.1 IntroductionThe Model 3010TAC Oxygen Analyzer Co

Strona 12 - Model 3010TAC Oxygen Analyzer

iiModel 3010TAC Teledyne Analytical InstrumentsCopyright © 1999 Teledyne Analytical InstrumentsAll Rights Reserved. No part of this manual may be repr

Strona 13 - Introduction

2-2: Part I2 Operational Theory Model 3010TAC Teledyne Analytical InstrumentsFigure 2-1: Block Diagram of the 3010TAC

Strona 14

Part I: 2-3Oxygen Analyzer Part I: Control Unit Teledyne Analytical InstrumentsThe digital concentration signal—along with input from the controlpane

Strona 15 - 1.4 Control Unit Rear Panel

2-4: Part I2 Operational Theory Model 3010TAC Teledyne Analytical Instruments

Strona 16

Oxygen Analyzer Part I: Control UnitPart I: 3-1 Teledyne Analytical InstrumentsInstallationInstallation of Model 3010TAC Analyzers includes:1. Unpack

Strona 17

3 Installation Model 3010TAC3-2: Part I Teledyne Analytical InstrumentsFigure 3-1: Front Panel of the Model 3010TAC Control UnitFigure 3-2: Single

Strona 18 - 1-6: Part I

Oxygen Analyzer Part I: Control UnitPart I: 3-3 Teledyne Analytical Instrumentsis pressed all the way in with a narrow gauge tool (less than 0.18 inc

Strona 19 - 2.1 Introduction

3 Installation Model 3010TAC3-4: Part I Teledyne Analytical InstrumentsCAUTION: Use Shielded Cables. Also, use plugs that provideexcellent EMI/RFI p

Strona 20 - 2-2: Part I

Oxygen Analyzer Part I: Control UnitPart I: 3-5 Teledyne Analytical InstrumentsAnalog Outputs: There are four DC output signal pins—two pinsper outp

Strona 21

3 Installation Model 3010TAC3-6: Part I Teledyne Analytical Instruments• Can be configured as failsafe or nonfailsafe.• Can be configured as latch

Strona 22 - 2-4: Part I

Oxygen Analyzer Part I: Control UnitPart I: 3-7 Teledyne Analytical Instrumentssynchronous signal must open and close external span valveappropriatel

Strona 23 - Installation

iii Oxygen Analyzer Teledyne Analytical InstrumentsTable of ContentsSpecific Model Information... ivPreface ...

Strona 24

3 Installation Model 3010TAC3-8: Part I Teledyne Analytical Instrumentssignals to ensure that the zero and span gas valves will becontrolled synchro

Strona 25 - 3.3 Electrical Connections

Oxygen Analyzer Part I: Control UnitPart I: 3-9 Teledyne Analytical InstrumentsThe voltage from these outputs is nominally 0 V for the OFF and15 V dc

Strona 26

3 Installation Model 3010TAC3-10: Part I Teledyne Analytical InstrumentsTable 3-5: Commands via RS-232 InputCommand Descriptionas<enter> Immed

Strona 27

Oxygen Analyzer Part I: Control UnitPart I: 3-11 Teledyne Analytical InstrumentsCAUTION: If you use your own control valves, or any otherdesign, plea

Strona 28

3 Installation Model 3010TAC3-12: Part I Teledyne Analytical Instruments

Strona 29

Oxygen Analyzer Part I: Control UnitPart I: 4-1 Teledyne Analytical InstrumentsOperation4.1 IntroductionOnce the analyzer has been installed, configu

Strona 30

4 Operation Model 3010TAC4-2: Part I Teledyne Analytical Instruments4.2 Using the Data Entry and FunctionButtonsData Entry Buttons: The < > ar

Strona 31

Oxygen Analyzer Part I: Control UnitPart I: 4-3 Teledyne Analytical Instruments4.3 The System FunctionThe subfunctions of the System function are de

Strona 32

4 Operation Model 3010TAC4-4: Part I Teledyne Analytical InstrumentsAfter a password is assigned, the operator must log out toactivate it. Until the

Strona 33 - 3.4 Testing the System

Oxygen Analyzer Part I: Control UnitPart I: 4-5 Teledyne Analytical InstrumentsIn the first line, TRACK or HOLD should be blinking. The operatorcan t

Strona 34 - 3-12: Part I

ivModel 3010TAC Teledyne Analytical InstrumentsSpecific Model InformationThe instrument for which this manual was supplied may incorporate oneor more

Strona 35 - Operation

4 Operation Model 3010TAC4-6: Part I Teledyne Analytical InstrumentsSpan OFF Nxt: 0d 0hZero OFF Nxt: 0d 0hPress < > arrows to blink Span (or

Strona 36

Oxygen Analyzer Part I: Control UnitPart I: 4-7 Teledyne Analytical InstrumentsTRAK/HLD Auto—CalPSWD Logout MoreUse the < > arrow keys to scrol

Strona 37 - 4.3 The System Function

4 Operation Model 3010TAC4-8: Part I Teledyne Analytical InstrumentsIf you chose Enter to change the password, the password assignmentscreen appears

Strona 38

Oxygen Analyzer Part I: Control UnitPart I: 4-9 Teledyne Analytical Instruments0.0 ppm AnlZRange: 0 — 100If an alarm is tripped, the second

Strona 39

4 Operation Model 3010TAC4-10: Part I Teledyne Analytical InstrumentsTRAK/HLD Auto—CalPSWD Logout MoreUse the < > arrow keys to blink More, th

Strona 40

Oxygen Analyzer Part I: Control UnitPart I: 4-11 Teledyne Analytical Instruments- Use the Right or Left arrow keys and select More. Press Enter.Vers

Strona 41

4 Operation Model 3010TAC4-12: Part I Teledyne Analytical InstrumentsIf you are using password protection, you will need to enter yourpassword to ga

Strona 42

Oxygen Analyzer Part I: Control UnitPart I: 4-13 Teledyne Analytical Instruments4.4.1.2 Manual Mode ZeroingPress Zero to enter the Zero function. The

Strona 43

4 Operation Model 3010TAC4-14: Part I Teledyne Analytical Instruments4.4.2 Span CalThe Span button on the front panel is used to span calibrate the

Strona 44

Oxygen Analyzer Part I: Control UnitPart I: 4-15 Teledyne Analytical Instrumentssponds, within tolerance, to the value of the span gas concentration.

Strona 45

v Oxygen Analyzer Teledyne Analytical InstrumentsPrefaceOverviewThe Teledyne Analytical Instruments Model 3010TAC Trace OxygenAnalyzer is a versat

Strona 46

4 Operation Model 3010TAC4-16: Part I Teledyne Analytical Instrumentsthe full scale of the range being calibrated for a period of ten minutes it iss

Strona 47

Oxygen Analyzer Part I: Control UnitPart I: 4-17 Teledyne Analytical InstrumentsDecide how your alarms should be configured. The choice will dependup

Strona 48

4 Operation Model 3010TAC4-18: Part I Teledyne Analytical InstrumentsSet up alarm 1 by moving the blinking over to AL–1 using the < >arrow key

Strona 49

Oxygen Analyzer Part I: Control UnitPart I: 4-19 Teledyne Analytical Instrumentsrange limits are exceeded, the instrument will automatically shift to

Strona 50 - 4.5 The Alarms Function

4 Operation Model 3010TAC4-20: Part I Teledyne Analytical InstrumentsTo switch from autoranging to fixed range analysis, enter the rangefunction by

Strona 51 - Choose Alarm

Oxygen Analyzer Part I: Control UnitPart I: 4-21 Teledyne Analytical Instrumentsconcentration and one range ID) and two isolated 4-20 mA dc currentou

Strona 52 - 4.6 The Range Function

4 Operation Model 3010TAC4-22: Part I Teledyne Analytical InstrumentsIMPORTANT:In the event of loss of flow through the analyzer, if thevent is vent

Strona 53

Part I: Control Unit Maintenance 5Part I: 5-1 Teledyne Analytical InstrumentsMaintenanceAside from normal cleaning and checking for leaks at the ga

Strona 54 - 4.8 Signal Output

5 Maintenance Model 3010TAC Oxygen Analyzer5-2: Part I Teledyne Analytical Instruments3. Replace fuse as shown in Figure 5-2.4. Reassemble Housing

Strona 55

Part I: Control Unit Maintenance 5Part I: 5-3 Teledyne Analytical Instruments5.3 Major Internal ComponentsThe major components in the Control Unit

Strona 56 - 4-22: Part I

viModel 3010TAC Teledyne Analytical InstrumentsModel and Part Number DesignationsThe part numbers are the most specific identification. When using thi

Strona 57 - Maintenance

5 Maintenance Model 3010TAC Oxygen Analyzer5-4: Part I Teledyne Analytical InstrumentsxxxxxFigure 5-4: Rear-Panel ScrewsTo detach the rear panel, r

Strona 58 - Figure 5-2: Installing Fuses

Part II: i Part II: Analysis Unit Teledyne Analytical InstrumentsOPERATING INSTRUCTIONSModel 3010TACOxygen AnalyzerPart IIIntrinsic Safe Barriersand

Strona 59 - 5.3 Major Internal Components

ii: Part IIModel 3010TAC Oxygen Analyzer Teledyne Analytical InstrumentsTable of Contents1 Introduction1.1 Overview ...

Strona 60 - 5.4 Cleaning

Part II: 1-1Oxygen Analyzer Part II: Analysis Unit Teledyne Analytical InstrumentsIntroduction1.1 OverviewThe Analytical Instruments Model 3010TAC Re

Strona 61

1-2: Part II1 Introduction Model 3010TAC Teledyne Analytical InstrumentsCAUTION: Bypassing the barriers in any way nullifies theireffect, and cond

Strona 62

Part II: 2-1Oxygen Analyzers Part II: Analysis Units Teledyne Analytical InstrumentsOperational Theory2.1 IntroductionThe Analysis Unit is composed

Strona 63

2-2: Part II2 Operational Theory Model 3010TAC Teledyne Analytical Instrumentsbeing analyzed. The Micro-Fuel Cell is therefore a hybrid between abat

Strona 64 - 1.4 Cell Housing/Probe

Part II: 2-3Oxygen Analyzers Part II: Analysis Units Teledyne Analytical InstrumentsAt the top end of the cell is a diffusion membrane of Teflon, wh

Strona 65 - 2.2 Micro-Fuel Cell Sensor

2-4: Part II2 Operational Theory Model 3010TAC Teledyne Analytical InstrumentsThe overall reaction for the fuel cell is the SUM of the half reaction

Strona 66 - Figure 2-1: Micro-Fuel Cell

Part II: 2-5Oxygen Analyzers Part II: Analysis Units Teledyne Analytical InstrumentsFigure 2-3. Characteristic Input/Output Curve for a Micro-Fuel

Strona 67

vii Oxygen Analyzer Teledyne Analytical InstrumentsMain Features of the AnalyzerThe Model 3010TAC series Oxygen Analyzers are sophisticated yetsim

Strona 68

2-6: Part II2 Operational Theory Model 3010TAC Teledyne Analytical Instruments

Strona 69

Oxygen Analyzer Part II: Analysis UnitPart II: 3-1 Teledyne Analytical InstrumentsInstallationInstallation of the Model 3010TAC Analyzer includes:1.

Strona 70 - 2-6: Part II

3 Installation Model 3010TAC3-2: Part II Teledyne Analytical InstrumentsIf you use your own gas control valves, use the interconnect diagramin Figur

Strona 71

Oxygen Analyzer Part II: Analysis UnitPart II: 3-3 Teledyne Analytical Instruments3.5 Intrinsic Safety BarriersTwo dual-channels intrinsic safety bar

Strona 72

3 Installation Model 3010TAC3-4: Part II Teledyne Analytical InstrumentsD - SUB CONNECTOR’S DESCRIPTIONPIN # Description 3 + Range ID 4-20 mA 4 -

Strona 73 - 3.5 Intrinsic Safety Barriers

Oxygen Analyzer Part II: Analysis UnitPart II: 3-5 Teledyne Analytical Instruments 40 Calibration Contact 41 Calibration Contact 42 Alarm 2 NC Contac

Strona 74

3 Installation Model 3010TAC3-6: Part II Teledyne Analytical Instruments

Strona 75

Oxygen Analyzer Part II: Analysis UnitPart II: 4-1 Teledyne Analytical InstrumentsMaintenance4.1 Routine MaintenanceAside from normal cleaning and ch

Strona 76 - 3-6: Part II

4 Maintenance Model 3010TAC4-2: Part II Teledyne Analytical Instruments4.2.1 Installing a New Micro-Fuel CellIt is important to minimize the amou

Strona 77

Oxygen Analyzer Part II: Analysis UnitPart II: 4-3 Teledyne Analytical Instruments4.2.2 Cell WarrantyThe Micro-Fuel cell used in the standard Model

Strona 78

viiiModel 3010TAC Teledyne Analytical InstrumentsModel 3010TAC complies with all of the requirements of theCommonwealth of Europe (CE) for Radio Freq

Strona 79

4 Maintenance Model 3010TAC4-4: Part II Teledyne Analytical Instruments0OK1 DAC A (0–1 V Concentration)2 DAC B (0–1 V Range ID)3 Both FailedPreamp0

Strona 80

A-1Oxygen Analyzer Appendix Teledyne Analytical InstrumentsOPERATING INSTRUCTIONSModels 3010TACOxygen AnalyzersAppendixFlush Mount Control Unit, PN D6

Strona 81 - Models 3010TAC

A-2Appendix Model 3010TAC Teledyne Analytical InstrumentsContentsA-1 Model 3010TAC Specifications... A-3A-2 Rec

Strona 82 - Contents

A-3Oxygen Analyzer Appendix Teledyne Analytical InstrumentsAppendixA-1 Models 3010TAC SpecificationsPackaging: General Purpose Control Unit• Flush pa

Strona 83 - Appendix

A-4Appendix Model 3010TAC Teledyne Analytical InstrumentsPower:General Purpose Control UnitUniversal power supply 85-250 V ac,47-63 Hz.Operating Te

Strona 84

A-5Oxygen Analyzer Appendix Teledyne Analytical InstrumentsA-2 Recommended 2-Year Spare Parts ListQty Part Number Description1 C67821 Back Panel Boa

Strona 85 - Orders should be sent to:

A-6Appendix Model 3010TAC Teledyne Analytical InstrumentsA-3 Drawing ListD-66192A:Final Assembly/Outline Drawing, Control Unit, Trace OxygenD-66193

Strona 86 - A-3 Drawing List

A-7Oxygen Analyzer Appendix Teledyne Analytical InstrumentsMeltingPoint(°C)328360A-4 Material Safety Data SheetProduct Name: Micro-Fuel Cells and Supe

Strona 87

A-8Appendix Model 3010TAC Teledyne Analytical InstrumentsFlash Point: na Flammable Limits: na LEL: na UEL: naExtinguishing Media: Use extinguishing

Strona 88 - Section V – Reactivity Data

A-9Oxygen Analyzer Appendix Teledyne Analytical InstrumentsContact of electrolyte with skin or eyes will cause aburning sensation and/or feel soapy or

Strona 89 - Protective measures

Part I: i Part I: Control Unit Teledyne Analytical InstrumentsOPERATING INSTRUCTIONSModel 3010TACOxygen AnalyzerPart I: Control UnitFlush Panel Mount

Strona 90 - Section X – References

A-10Appendix Model 3010TAC Teledyne Analytical InstrumentsSection VIII – Control MeasuresEye Protection: Chemical splash gogglesHand Protection: Ru