MOLY TEST DATA SHEET
Line # TEST RESULT
2 LEAK-CHECK (WHEN HOT) YES / NO
3 NO
X
RESPONSE (MOLY BYPASSED) __________
4 NO
X
RESPONSE (MOLY IN-LINE) __________
5 OUT-GASSING / EATING (NO – NO
X
) __________ (>-5 PPB, <5
PPB)
6 (NOx
ORIG
) (NO
X
mode, O
3
off) __________ PPB
7 (NOx
REM
) (NO
X
mode, O
3
on) __________ PPB
8 NO
X
LOSS (9A - 10B) __________ (A) (<4% of
NOx
ORIG;
ex: for
450PPB
4% is 18PPB)
9 (NO
ORIG
) (NO mode, O
3
off) __________ PPB
10 (NO
REM
) (NO mode, O
3
on) __________ PPB
11 NO
2
(9B - 10A) __________ (B)
(>300PPB)
12 Efficiency LOSS [ (
A / B ) x 100 ] = [ ( ____A____ / ____B____ ) x 100 ] =
____C____%
13 Total Conv Eff [ 100% – C ] = 100% - ____C_____ = _____D_____ % ( > 96%)