MANUAL ADDENDUM
M200EU
NOy CONVERTER OPTION
Teledyne Advanced Pollution Instrumentation
9480 Carroll Park Dr.
San Diego, CA 92121-5201
TOLL-FREE: 800-324-5190
FAX: 858-657-9818
TEL: 858-657-9800
E-MAIL:
WEB SITE: www.teledyne-api.com
05386
REV. A
Copyright 2006 API Inc. 10 October 2006
Przeglądanie stron 0
1 2 3 4 5 6 ... 32 33

Podsumowanie treści

Strona 1 - NOy CONVERTER OPTION

MANUAL ADDENDUM M200EU NOy CONVERTER OPTION Teledyne Advanced Pollution Instrumentation 9480 Carroll Park Dr. San Diego, CA 92121-5

Strona 2 - SAFETY MESSAGES

Model 200EU-NOy Option Manual Addendum 2.0 Specifications, Warranty 2.1 Specifications Converter Remotely mounted molybdenum converter with tempera

Strona 3 - Table of Contents

Model 200EU-NOy Option Manual Addendum 2.2 Warranty ADVANCED POLLUTION INSTRUMENTATION, INC. 02024c Prior to shipment, API equipment is thoroughly

Strona 4

Model 200EU-NOy Option Manual Addendum 3.0 The M501Y NOy Converter 3.1 Principle of Operation The API Model 200EU with NOy Option is designed to me

Strona 5 - List of Figures

Model 200EU-NOy Option Manual Addendum 4.0 CALIBRATION AND ZERO/SPAN CHECKS Unlike most NOx analyzers, the M200EU with the NOy option does not have

Strona 6 - List of Tables

Model 200EU-NOy Option Manual Addendum 4.1 Calibration or Cal Check Procedure The calibration of the instrument can be checked or adjusted using

Strona 7 - 1.0 Getting Started

Model 200EU-NOy Option Manual Addendum Enter the expected NOx and NO span gas concentrations: Table 4.2 - Enter Expected Span Gas Concentrations Pr

Strona 8

Model 200EU-NOy Option Manual Addendum Table 4.3 - Span Calibration Procedure Step Number Action Comment 1. Press CAL The M200EU enters the cali

Strona 9 - 1.3 Initial Operation

Model 200EU-NOy Option Manual Addendum 05386 Rev. A 17 5.0 Maintenance 5.1 Maintenance Schedule See Table 5.1 below.

Strona 11 - 2.2 Warranty

Model 200EU-NOy Option Manual Addendum 05386 Rev. A 19 Table 5.1- Preventative Maintenance Calendar Item Maintenance Interval Zero/Span

Strona 12 - 3.0 The M501Y NOy Converter

Model 200EU-NOy Option Manual Addendum SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important saf

Strona 13

Model 200EU-NOy Option Manual Addendum 5.2 Replacing the Sample Particulate Filter The particulate filter should be inspected often for signs of plu

Strona 14

Model 200EU-NOy Option Manual Addendum Figure 5.1 - Replacing the Particulate Filter 05386 Rev. A 21

Strona 15

Model 200EU-NOy Option Manual Addendum 5.3 Check Bypass Flow Rates The external sample pump is capable of maintaining the cell pressure at less tha

Strona 16

Model 200EU-NOy Option Manual Addendum Figure 5.2 - Converter Assembly 05386 Rev. A 23

Strona 17 - 5.0 Maintenance

Model 200EU-NOy Option Manual Addendum 5.5 Pneumatic Line Inspection Particulate matter in the pneumatic lines will affect both flow rate and respon

Strona 18

Model 200EU-NOy Option Manual Addendum Figure 5.3- Pneumatics diagram (standard and Z/S option) 05386 Rev. A 25

Strona 19

Model 200EU-NOy Option Manual Addendum Figure 5.4: Pneumatic diagram (OPT 86 option) 5.6 Leak Check Procedure 05386 Rev. A 26

Strona 20

Model 200EU-NOy Option Manual Addendum If a leak checker is not available, it is possible to leak check the instrument using the M501Y's pump p

Strona 21 - 05386 Rev. A 21

Model 200EU-NOy Option Manual Addendum 6.0 Troubleshooting, Adjustments NOTE THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY QUALI

Strona 22 - 5.4 Replacing the Converter

Model 200EU-NOy Option Manual Addendum 6.1.3 Bypass Pump Diagnostic Procedures The bypass pump is capable of maintaining a total bypass flowrate of

Strona 23

Model 200EU-NOy Option Manual Addendum Table of Contents SAFETY MESSAGES Table of Contents List of Figures List of Tables 1.0 GETTING STARTED 1.1 Un

Strona 24

Model 200EU-NOy Option Manual Addendum 6.1.4 Temperature Controller Setup The temperature controller has been programmed at the factory and should

Strona 25

Model 200EU-NOy Option Manual Addendum SECONDARY MENU (The fixed characteristics of the system). PRESS DISPLAY ACTION SEL HOLD TILL p-n1 SET TO

Strona 26

Model 200EU-NOy Option Manual Addendum Figure 6.1 – Electrical Diagram 05386 Rev. A 32

Strona 27

Model 200EU-NOy Option Manual Addendum 7.0 M501Y Spare Parts List Note: Use of replacement parts other than those supplied by API may result in non

Strona 28 - 6.1.2 Leak Check

Model 200EU-NOy Option Manual Addendum 6.1 Operation Verification - M501Y Diagnostic Techniques 6.1.1 Pneumatic System 6.1.2 Leak Check 6.1.3 Bypass

Strona 29

Model 200EU-NOy Option Manual Addendum List of Figures FIGURE 1.1 - REAR PANEL PNEUMATIC CONNECTIONS ...

Strona 30 - PRESS DISPLAY ACTION

Model 200EU-NOy Option Manual Addendum 05386 Rev. A 6 List of Tables TABLE 2.1 - FINAL TEST AND CALIBRATION VALUES TABLE 4.1 - MANUAL ZERO CALIB

Strona 31

Model 200EU-NOy Option Manual Addendum 1.0 Getting Started The M501Y has been designed to operate with the M200EU low level NOx analyzer. 1.1 Unpa

Strona 32

Model 200EU-NOy Option Manual Addendum Figure 1.1 - Rear Panel Pneumatic Connections 05386 Rev. A 8

Strona 33 - 7.0 M501Y Spare Parts List

Model 200EU-NOy Option Manual Addendum 1.3 Initial Operation 1. After confirming proper supply voltage, turn on the instrument power. If you are