Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E
2-4
2.7 Options
The M452 is available with the following options:
2.7.1 AC Power Adapter
This option consists of a Universal Input (100-240 VAC, 50-60Hz) Power
Adapter with a standard IEC320 modular AC receptacle. A North
American standard 3 prong power cord is also provided.
2.7.2 Ozone Destructor (1/4” Tube Fittings)
This option consists of an external catalytic Ozone Destructor for
connecting to the gas outlet of the M452. The Ozone Destructor is
provided with ¼” compression tube fittings. When the Ozone Descructor
is used in conjunction with the M452, the gas flow rate must be limited to
a maximum of 1.0 L/min.