Teledyne T101 Instrukcja Użytkownika Strona 169

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 168
Teledyne API – T101 Operation Manual Calibration Procedures
167
STEP TWO: Set the expected H
2
S span gas concentrations. In this example the
instrument is set for single (SNGL) range mode with a reporting range span of 500 ppb.
This sequence causes the
analyzer to pro mpt for the
expected H
2
S spa n
concentration .
M-P CAL H2S SPAN CONC: 450.0 Con
c
0 0 0 4 5 0 .0 ENTR EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > ZERO CONC EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
The H
2
S span concentration
values automatically de faul t
to 450.0 ppb.
To change this value to the
actual concentration of the
sp an g as, ente r the nu mbe r
by pressing the key under
each dig it until the e xpected
value a ppear s.
The span gas concentration
should always be 90% of the
se lected rep ortin g rang e
EXAMPLE
Rep orting range = 800 ppb
Span gas conc.= 720 ppb
EXIT ignores th e ne w se ttin g
and r etur ns to the pr evio us
display.
ENTR accepts the new setting
and returns to the
pre vious disp lay..
07266B DCN6485
Przeglądanie stron 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 370 371

Komentarze do niniejszej Instrukcji

Brak uwag