Teledyne T101 Instrukcja Użytkownika Strona 97

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 96
Teledyne API – T101 Operation Manual Operating Instructions
95
4.6.3.6. AIN Calibration
This is the sub-menu to conduct the analog input calibration. This calibration should only
be necessary after major repair such as a replacement of CPU, motherboard or power
supplies. Activate the ANALOG I/O CONFIGURATION MENU (see Section 4.6.3),
then press:
Exit at any tim e to
r eturn to th e mai n
DIAG menu
DIAG
A
NALOG I / O CONFIGURATION
PR EV NEXT ENTR EXIT
Contin ue pressing SET> until
DIAG AIO
A
IN CALIBRATED: NO
< SET SET> CAL EXIT
DIAG AIO
A
IN CA LIBRA TED: YES
< SET SET> CAL EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG AIO CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
07266B DCN6485
Przeglądanie stron 96
1 2 ... 92 93 94 95 96 97 98 99 100 101 102 ... 370 371

Komentarze do niniejszej Instrukcji

Brak uwag